Calibrate Resistance Level

Perform this calibration procedure to display the correct resistance levels throughout the full range of the resistance lever. This is required when installing the unit, replacing the console or sensor board, or if the unit has been disassembled.
  1. Press and hold the Up arrow and Down arrow for 3 seconds.
  2. Navigate in the Toolbox menu to DIAGNOSTICS with the Down arrow.
  3. Press the GO/Enter key to enter DIAGNOSTICS.


  4. Push the resistance lever forwards to the highest resistance level.
  5. Press and hold the Up arrow for three seconds until the BRAKE VALUE section blinks. The upper value is stored.
  6. Pull the resistance lever backwards to the lowest resistance level.
  7. Press and hold the Down arrow for three seconds until the BRAKE VALUE section blinks. The lower value is stored.
  8. Press the GO/Enter key to go BACK to the toolbox menu.
  9. Press the Down arrow to select EXIT in the setup menu.
  10. Press the GO/Enter key to exit Toolbox.
Table 1. Diagnostics menu
ARC TRAINER Software revision number.
PRESSURE/TEMP

Pressure: Displays Barometric pressure in HPA (hectopascal).

Temperature : Displays local temperature in degrees Celsius.

BRAKE VALUE Displays brake position value. Range is P1 to P10.
BATTERY Displays battery level from 0 to 100% in 10% increments. LOW BATTERY screen will be displayed at 10%.